| MEASURING
RANGES |
Carbon
Up to 500mg C resp. up to 100% C at 500 mg sample1) |
Hydrogen
Up to 75mg H resp. up to 15% H at 500 mg sample1)
|
Sulfur
Up to 10mg S resp. up to 2% S at 500 mg sample1)
|
Indicating
range
Up to 100% C2) |
Indicating range
Up to 100% H2) |
Indicating range
Up to 100% S2) |
| SENSITIVITY |
Carbon
5 µg C resp. 10 ppm C at 500 mg sample |
Hydrogen
50 µg H resp. 100 ppm H at 500 mg sample |
Sulfur
1 µg S resp. 2 ppm S at 500 mg sample |
| ACCURACY |
Carbon
±50 µg C resp. ±100ppm C at 500 mg sample or ± 1% of C
present1) |
Hydrogen
±100µg H resp. ±200ppm H at 500 mg sample or ± 1% of H
present1) |
Sulfur
±2µg S resp. ±4ppm S at 500 mg sample or ± 1% of S
present1) |
| GENERAL SPECIFICATIONS |
Normal
sample weight
200-300 mg for coal |
Normal analysis time
50 to 180 seconds |
Furnace
temperature
Adjustable up to 1550 °C |
Gas required
Oxygen 2 to 4 bar (30 to 60 psi) 3l/min |
Interfaces
Computer - serial4) |
Chemicals
CO2 trap - sodium hydroxide
H2O trap - magnesium perchlorate |
Analyzer
power requirements
230 V AC ±10% 50/60 Hz
0.33 A / 75 W |
Furnace power requirements
230 V AC ±10% 50/60 Hz
Maximum heat up current 20 A |
Weights
Analyzer Furnace
40kg 36kg
|
Dimensions
Width Height Depth
Analyzer: 55cm (21") 52cm (20.5")
60cm (23.5") Furnace: 33cm
(13") 52cm (20.5") 60cm (23.5")3)
|
| ACCESSORIES |
Balance:
0.0001g to 60 g ± 0.0001 g |
|
Computer: PC with HDD, 3.5" drive, CD-ROM, TFT flat screen
and keyboard |
Color
printer with automatic cut sheet feed, other options
on request |