| MEASURING RANGES |
Low carbon
Up to 25mg C resp. up to 5%C at 500mg sample 1) |
Low sulfur
Up to 10mg S resp. up to 2% S at 500mg sample 1)
|
High carbon
Up to 500mg C resp. up to 100%C at 500mg sample1) |
High sulfur
Up to 100mg S resp. up to 100% S at 100mg sample1) |
| SENSITIVITY |
Carbon
5µg C resp. 10 ppm C at 500mg sample 1) |
Sulfur
1µg S resp. 2 ppm S at 500mg sample 1) |
| ACCURACY |
Low
carbon 1)
±10µg C resp. ±20ppm C at 500mg sample or ±1% of C present
|
Low sulfur
1)
±2µg S resp. ±4ppm S at 500mg sample or ±1% S present |
High
carbon 1)
±50µg C resp. ±100ppm C at 500mg sample or ±1% of C
present |
High sulfur 1)
±50µg S resp. ±100ppm S at 500mg2) sample or ±1% S present |
| GENERAL SPECIFICATIONS |
Normal
sample weight
400mg for coal |
Normal analysis time
60 to 120 sec. |
Furnace temperature
Adjustable up to 1550 C° |
Gas required
Oxygen 99.5% pure 2 to 4 bar (30 to 60 psi) 3 l/min
|
Interfaces
serial or USB 4) |
Chemicals
CO2 Trap - Sodium hydroxide
H2O Trap - Magnesium perchlorate |
Detection method
Solid state infrared absorption for CO2 and SO2 |
Analyser power requirements
230 V AC ±10% 50/60 Hz
Maximum heat-up current 20 A
|
Weight
Analyzer TIC-Mudule
60kg 28kg
|
Dimensions (Width x Height x Depth)
Analyzer - 55cm(21") x 52cm(20.5") x 60cm(23.5")
TIC-Module - 33cm(13") x 52cm(20.5")
x 60cm(23.5") 3) |
| ACCESSORIES |
Balance:
0.0001g to 60 g ±0.0001 g |
Computer:
PC with HDD, 3.5" drive, CD-ROM, TFT flat screen
and keyboard |
Color
printer with automatic cut sheet feed, other options
on request |