CS-530
Carbon / Sulfur Determinator

  Specifications
MEASURING RANGES

Low carbon
Up to 25mg C resp. up to 5%C at 500mg sample 1)

Low sulfur
Up to 10mg S resp. up to 2% S at 500mg sample 1)

High carbon
Up to 500mg C resp. up to 100%C at 500mg sample1)

High sulfur
Up to 100mg S resp. up to 100% S at 100mg sample1)

SENSITIVITY

Carbon
5µg C resp. 10 ppm C at 500mg sample 1)

Sulfur
1µg S resp. 2 ppm S at 500mg sample 1)

ACCURACY

Low carbon 1)
±10µg C resp. ±20ppm C at 500mg sample or ±1% of C present

Low sulfur 1)
±2µg S resp. ±4ppm S at 500mg sample or ±1% S present

High carbon 1)
±50µg C resp. ±100ppm C at 500mg sample or ±1% of C present

High sulfur 1)
±50µg S resp. ±100ppm S at 500mg2) sample or ±1% S present

GENERAL SPECIFICATIONS

Normal sample weight
400mg for coal

Normal analysis time
60 to 120 sec.

Furnace temperature
Adjustable up to 1550 C°

Gas required
Oxygen 99.5% pure 2 to 4 bar (30 to 60 psi) 3 l/min

Interfaces
serial or USB 4)

Chemicals
CO2 Trap - Sodium hydroxide
H2O Trap - Magnesium perchlorate

Detection method
Solid state infrared absorption for CO2 and SO2

Analyser power requirements
230 V AC ±10% 50/60 Hz
Maximum heat-up current 20 A

Weight
Analyzer   TIC-Mudule
60kg        28kg

Dimensions (Width x Height x Depth)
Analyzer - 55cm(21") x 52cm(20.5") x 60cm(23.5")
TIC-Module - 33cm(13") x 52cm(20.5") x 60cm(23.5") 3)

ACCESSORIES

Balance: 0.0001g to 60 g ±0.0001 g

Computer: PC with HDD, 3.5" drive, CD-ROM, TFT flat screen and keyboard

Color printer with automatic cut sheet feed, other options on request

  1) Other ranges on request.  2) Reduce the sample weight for samples over 20% S.  3) Allow 15 cm (6”) access area behind the furnace. 4) Balance (serial - RS232) and printer (USB, parallel) are connected to the PC.