| MEASURING RANGES |
Low
carbon
Up to 0.1% C at 500mg sample resp. up to 0.5mg C1) |
Low sulfur
Up to 0.3% S at 500mg sample resp. up to 1.5mg S1)
|
High
carbon
Up to 5% C at 500mg sample resp. up to 25mg C 1)
Indicating range Up to 100% C2) |
High sulfur
30% S at 150mg sample resp. up to 45mg S1) Indicating range up to 100% S2) |
| SENSITIVITY |
Carbon
0.1 ppm C at 500mg sample resp. 0.05µg C |
Sulfur
0.1 ppm S at 500mg sample resp. 0,05µg S 1) |
| ACCURACY |
Low
carbon 1)
±1ppm C
3) at 1gram sample resp. ±1 µg
C or ±0.5% of carbon present |
Low sulfur
1)
±1ppm S at 1gram sample resp. ±1µg S or ±0.5% of sulfur present |
High
carbon 1)
±100ppm C
3) at 500 mg sample resp. ±50 µg C or ±0.5% of C present |
High sulfur 1)
±0.1% S at 150 mg sample resp. ±150 µg
S or ±0.5% of S present |
| GENERAL SPECIFICATIONS |
Normal
sample weight
0.5g to 1g for steel and cast iron |
Normal analysis time
40 to 50 seconds |
Furnace
type
Induction, 19.5 MHz 2.2 kVA max |
Furnace dust cleaning
Automatic |
Detection
method
Solid state infrared absorption
for carbon and sulfur |
Chemicals
CO2 Trap Sodium hydroxide
H2O trap Magnesium perchlorate Catalyst copper oxide |
Gas required
Oxygen 99.5% pure, 2 to 4 bar (30 to 60 psi), 3 l/min |
Compressed air
4 to 6 bar (60 to 90 psi) |
Interfaces
Computer - serial 4) |
Power requirements
230 V AC ±10% 50/60 Hz max 15 Amps 3450 Watts |
Weight
Analyzer: approx. 110 kg
|
Dimensions
Width
Height Depth
55cm(21") 80cm(31.5") 60cm(23.5")
|
| ACCESSORIES |
Balance:
0.0001g to 60 g ±0.0001 |
Computer:
Pentium PC with HDD, 3.5" drive, CD-ROM, color
monitor and keyboard |