| MEASURING
RANGES |
CO2
Up to 400mg CO2
1) |
H2O
Up to 300mg H2O
1)
|
Indicating
range
Up to 100% CO2
2) |
Indicating range
Up to 100% H2O 2) |
| SENSITIVITY |
CO2
0.0001% CO2 |
H2O
0.0001% H2O |
| ACCURACY |
CO2
± 0.02% CO2 or ± 1% of CO2
present 1) |
H2O
± 0.02% H2O or ± 1% of H2O
present 1) |
| GENERAL SPECIFICATIONS |
Detection
method
Solid state infrared absorption for CO2 and H2O |
Normal sample weight
200 mg |
Furnace
temperature
Adjustable up to 1000 °C |
Normal analysis time
2 to 3 minutes |
Interfaces
Computer - serial 4) |
Gas required
Nitrogen 2 to 4 bar (30 to 60 psi) 3 l/min |
Power
requirements
230 V AC ±10% 50/60 Hz
10 A / 2.3 kW |
Chemicals
CO2 trap sodium hydroxide
H2O trap magnesium perchlorate |
Weight
Analyzer
approx. 65 kg |
Dimensions
Width Height
Depth
55cm (21") 80cm (31.5") 60cm (23.5")
|
| ACCESSORIES |
Balance:
0.0001g to 60 g ± 0.0001 g |
Computer:
PC with HDD, 3.5" drive, CD-ROM, TFT flat screen
and keyboard |
Color
printer with automatic cut sheet feed, other options
on request |