| MEASURING
RANGES |
| Hydrogen
at 5 gram sample |
| Low
hydrogen
0 - 50 ppm H |
High hydrogen
Up to 1% H |
| Sensitivity
0.01 ppm H |
| ACCURACY |
| ±
0.05 ppm H or ± 2.5% of the measured value, depending
on which value is greater
|
| GENERAL SPECIFICATIONS |
Max.
sample weight
10 g |
Max. sample diameter
8 mm |
Max.
sample length
60 mm |
Detection method
Thermal conductivity |
Interfaces
Computer - serial 1) |
Carrier gas required
Nitrogen min 99.995% pure, 2 to 4 bar (30 to 60 psi),
30 l/h |
Power
requirements
230 VAC ±10% 50/60 Hz
2.0 Amps; 450 Watts |
Chemicals
CO2 trap sodium hydroxide
H2O trap magnesium perchlorate
Schutzes reagent |
Weights
approx. 40 kg |
Dimensions
Width Height
Depth
75cm
(29.5") 52cm (20.5") 60cm (23.5")
|
| ACCESSORIES |
Balance:
0.0001g to 60 g ± 0.0001 g |
Computer:
PC with HDD, 3.5" drive, CD-ROM, TFT flat screen
and keyboard |
Color
printer with automatic cut sheet feed, other options