| MEASURING
RANGES |
Oxygen
at 1 gram sample |
Hydrogen
at 1 gram sample |
Low oxygen
0.0 - 300 ppm 1) |
High oxygen
up to 2% 1) |
Low hydrogen
0.0 - 50 ppm 1) |
High hydrogen
up to 1000 ppm 1) |
Sensitivity 0.01 ppm oxygen
Indicating range 0 - 100% oxygen 2) |
Sensitivity
0.01 ppm hydrogen
Indicating range 0 - 100% hydrogen
2) |
| ACCURACY |
Low oxygen
± 0.1 ppm or ± 1% of oxygen present |
Low hydrogen
± 0.05 ppm or ± 1% of hydrogen present |
High
oxygen
± 2 ppm or ± 1% of oxygen present |
High
hydrogen
± 0.5 ppm or ± 1% of hydrogen present |
| GENERAL SPECIFICATIONS |
Normal
sample weight
1g |
Normal
analysis time
2 minutes |
Detection
method
Solid state infrared absorption for oxygen
Thermal conductivity for hydrogen |
Chemicals
CO2 trap - sodium hydroxide
H2O trap - magnesium perchlorate
Catalyst - Schutzes reagent |
Gas
required
Nitrogen at least 99.995% pure
3) 2 to 4 bar (30 to 60 psi) |
Compressed
air
4 to 6 bar (60 to 90 psi) |
Calibration
Standard samples and gas dosing device on request |
Furnace temperature up to 3000 °C
Furnace power 0 to 8 kW |
Interfaces
Computer - serial 4) |
Power requirements max 8500 W 50/60 Hz 400VAC ±10%
3 phase + neutral or 230VAC ±10% single phase on request |
Weights
approximately 135 kg |
Dimensions
Width Height
Depth
55cm
(21") 80cm (31.5") 60cm (23.5")
|
| ACCESSORIES |
Balance:
0.0001g to 60 g ± 0.0001 g |
Computer:
Pentium PC with HDD, 3.5" drive, CD-ROM, TFT flat
screen and keyboard |
Color
printer with automatic cut sheet feed, other options
on request |