INFORMATION SPECIFICATIONS  

ON-900
Oxygen / Nitrogen Determinator

  Specifications
MEASURING RANGES

Oxygen at 1 gram sample

Nitrogen at 1 gram sample

Low oxygen
0.0 - 300 ppm1)

High oxygen
up to 2% 1)

Low nitrogen
0.0 - 300 ppm1)

High nitrogen
up to 2%1)

Sensitivity 0.01 ppm oxygen
Indicating range 0 - 100% oxygen2)

Sensitivity 0.1 ppm nitrogen
Indicating range 0 - 100% nitrogen2)

ACCURACY

Low oxygen
± 0.1 ppm or ± 1% of oxygen present

Low nitrogen
± 0.1 ppm or ± 1% of nitrogen present

High oxygen
± 2 ppm or ± 1% of oxygen present

High nitrogen
± 2 ppm or ± 1% of nitrogen present

GENERAL SPECIFICATIONS

Normal sample weight
1g

Normal analysis time
2 minutes

Detection method
Solid state infrared absorption for oxygen
Thermal conductivity for nitrogen

Chemicals
CO2 trap - sodium hydroxide
H2O trap - magnesium perchlorate
Catalyst - copper oxide

Gas required
Helium3) at least 99.995% pure4) 2 to 4 bar (30 to 60 psi)

Compressed air
4 to 6 bar (60 to 90 psi)

Calibration
Standard samples and gas dosing device on request

Furnace temperature up to 3000 °C
Furnace power 0 to 8 kW

Interfaces
Computer - serial5)

Power requirements
max 8500 W 50/60 Hz 400VAC ±10% 3 phase + neutral or 230VAC ±10% single phase on request

Weights
approximately 135 kg

 Dimensions
         Width         Height          Depth        
  55cm (21") 80cm (31.5") 60cm (23.5")   

ACCESSORIES

Balance: 0.0001g to 60 g ± 0.0001 g

Computer: Pentium PC with HDD, 3.5" drive, CD-ROM, color monitor and keyboard

Color printer with automatic cut sheet feed, other options on request

  1) Other ranges on request.  2) Possible by reducing the sample weight.  3) For oxygen analysis only, nitrogen can be used as carrier gas.  4) 99.999% pure for low soxygen analysis. 5) Balance (serial - RS232) and printer (USB, parallel) are connected to the PC.