| MEASURING
RANGES |
Low
0.1 to 100 µg/cm2 1) |
High
0.1 to 1000 µg/cm2 1)
|
| Resolution
0.1 µg/cm2
|
| ACCURACY |
Low
0.2 µg/cm2 or ± 1% of the demand value |
High
1 µg/cm2 or ± 1% of the demand value |
| GENERAL SPECIFICATIONS |
Furnace
sample diameter
32 mm max. |
Carrier gas pressure
2 to 4 bar
Gas flow approx.180 l/h |
Furnace
sample length
145 mm max. |
Normal analysis time
approx. 60-90 sec. |
Furnace
temperature
Adjustable up to 1000 °C |
Carrier gas
Oxygen 99.5% |
Furnace
resolution
1°C |
Chemicals
Magnesium perchlorate (Anhydrone) Sodium hydroxide (Ascarite) / Dust trap quartz wool |
Furnace
display
Digital |
Analyzer power requirements
230V 50/60Hz 1A
Furnace power requirements 220V 50/60Hz
10A max.(during warming-up time) |
Weights
Analyzer Furnace
26.5 kg 16
kg |
Dimensions
Width Height Depth
Analyzer: 50cm (19") 58cm
(22") 50cm (19") Furnace: 30cm
(11.5") 47cm (18") 50cm (19")
|
| ACCESSORIES |
Computer:
Pentium PC with HDD, 3.5" drive, CD-ROM, color
monitor and keyboard |
Color
printer with automatic cut sheet feed, other options
on request |